  • Title: [Micromethod for the determination of glucose with glucose dehydrogenase in the autoanalyzer (author's transl)].
    Author: Scholer A, Pianezzi A.
    Journal: J Clin Chem Clin Biochem; 1976 Apr; 14(4):189-95. PubMed ID: 1271021.
    Abstract:
    A new AutoAnalyzer method is described for the determination of glucose in 20 mul of haemolysate, urine, or cerebrospinal fluid, using glucose dehydrogenase. Without loss of precision, accuracy, sensitivity, or linearity, 900 samples may be analyzed, using the same volume of reagents normally required for the manual analysis of 75 samples. A comparison with the hexokinase method yields a correlation of 0.9978. A haemolysing solution compatible with the reagents used is described.
    [Abstract] [Full Text] [Related] [New Search]