These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: A microdialysis technique for continuous subcutaneous glucose monitoring in diabetic patients (part 2).
    Author: Varalli M, Marelli G, Maran A, Bistoni S, Luzzana M, Cremonesi P, Caramenti G, Valgimigli F, Poscia A.
    Journal: Biosens Bioelectron; 2003 Jul; 18(7):899-905. PubMed ID: 12713913.
    Abstract:
    The aim of this study was to evaluate the reproducibility, the accuracy and the reliability of a continuous subcutaneous glucose measuring system. The GlucoDay system (A. Menarini I.F.R. S.r.l.-Florence, Italy) is a portable instrument provided with a micro-pump and a biosensor, coupled to a microdialysis system (see part 1). This instrument has demonstrated high reliability coupled with a low degree of invasivity. The profiles of glucose monitoring allow to achieve an excellent knowledge of the real variation of glucose in diabetic patients. The reproducibility study showed a bias lower than 10% between instruments. The accuracy study showed a difference from the reference method lower than 15%.
    [Abstract] [Full Text] [Related] [New Search]