These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Immobilized glucose oxidase used in the continuous-flow determination of serum glucose. Author: León LP, Narayanan S, Dellenbach R, Horvath C. Journal: Clin Chem; 1976 Jul; 22(7):1017-23. PubMed ID: 1277498. Abstract: We used a tubular glucose-oxidase wall reactor in the "AutoAnalyzer II" continuous-flow analytical system to determine glucose in blood serum. Sensitivity was high and wash characteristics were satisfactory with use of a 30-cm tube containing immobilized glucose oxidase. Results compared favorably with those of the conventional free-enzyme method. More than 25000 such assays can be performed with a single enzyme tube, which also shows long-term storage stability. Because of the steady-state chemistry 60 samples can be analyzed per hour. The linearity of the method is excellent and sample interaction from 5.0 to 1.0 g/liter is less than 5%. Results correlate well (greater than 0.993) with those obtained with both the neocuproine method used in the "SMA 12/60" multichannel analytical system and the free glucose oxidase method used in the AutoAnalyzer.[Abstract] [Full Text] [Related] [New Search]