Title: Defibrillator-embedded rapid recovery electrocardiogram amplifier. Author: Neycheva T, Krasteva V. Journal: J Med Eng Technol; 2003; 27(4):178-85. PubMed ID: 12851062. Abstract: One of the most important performances of the defibrillator-embedded amplifier-monitor-recorder tract, connected to defibrillator electrodes, is its rapid recovery after the application of the shock pulse. Practically near-immediate restoration of the signal trace is mandatory for studies of post-shock effects on the myocardium. Automatic analysis of the electrocardiogram signal in public-access defibrillation, aiming for about 100% correct recognition of shockable and non-shockable rhythms, now requires fast amplifier settling, as the decision time should not exceed 10-20 s. Two circuits of post-shock amplifier transient suppressors were developed with non-linear feedback, resulting in second-order high-pass filtering, with gradual return to normally accepted first-order response. Simulation and testing in real conditions resulted in recovery periods in the range of 1-2 s for an amplifier tract of 1-30 Hz bandwidth, depending on the pulse waveform and electrode type.[Abstract] [Full Text] [Related] [New Search]