These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Defibrillator-embedded rapid recovery electrocardiogram amplifier.
    Author: Neycheva T, Krasteva V.
    Journal: J Med Eng Technol; 2003; 27(4):178-85. PubMed ID: 12851062.
    Abstract:
    One of the most important performances of the defibrillator-embedded amplifier-monitor-recorder tract, connected to defibrillator electrodes, is its rapid recovery after the application of the shock pulse. Practically near-immediate restoration of the signal trace is mandatory for studies of post-shock effects on the myocardium. Automatic analysis of the electrocardiogram signal in public-access defibrillation, aiming for about 100% correct recognition of shockable and non-shockable rhythms, now requires fast amplifier settling, as the decision time should not exceed 10-20 s. Two circuits of post-shock amplifier transient suppressors were developed with non-linear feedback, resulting in second-order high-pass filtering, with gradual return to normally accepted first-order response. Simulation and testing in real conditions resulted in recovery periods in the range of 1-2 s for an amplifier tract of 1-30 Hz bandwidth, depending on the pulse waveform and electrode type.
    [Abstract] [Full Text] [Related] [New Search]