These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Loss of carbon dioxide from serum samples exposed to air. Effect on blood gas parameters and strong ions. Author: Kirschbaum B. Journal: Clin Chim Acta; 2003 Aug; 334(1-2):241-4. PubMed ID: 12867298. Abstract: BACKGROUND: The acid-base status of patients is frequently inferred from the concentration of total carbon dioxide in a venous blood sample. Prior studies of hemodialysis patients have reported lower values for tCO(2) when blood is shipped to a distant central laboratory compared to samples assayed at a local laboratory. Loss of tCO(2) from the sample after exposure of the blood to room air was postulated as the predominant factor. METHODS: Serum samples were assayed on-site before and at varying intervals after removing the tube stopper on an AVL OPTI Critical care Analyzer using E-Cl cuvettes. RESULTS: A rapid and progressive decline in pCO(2) and increase in pH was documented over a 2 h period. However, the change in tCO(2) during this time period was small and not significant. There were no changes in the concentrations of strong ions. CONCLUSIONS: The discrepancy in tCO(2) reported previously cannot be attributed solely to rapid loss of CO(2) unless assay conditions greatly enhance the rate of loss compared to the experimental conditions used in this study.[Abstract] [Full Text] [Related] [New Search]