These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Spatial resolution of numerical models of man and calculated specific absorption rate using the FDTD method: a study at 64 MHz in a magnetic resonance imaging coil.
    Author: Collins CM, Smith MB.
    Journal: J Magn Reson Imaging; 2003 Sep; 18(3):383-8. PubMed ID: 12938138.
    Abstract:
    PURPOSE: To examine how fine a model resolution is necessary for calculation of specific energy absorption rate (SAR) for the human head in regions as small as 1 g. MATERIALS AND METHODS: Here we perform a simple study comparing the maximum SAR averaged over any 1 cm(3) and SAR averaged over the entire head for several models of the same human head within the same radiofrequency coil, but with spatial resolutions varying from 8-100 Yee cells per cm(3). RESULTS: Over the range of model resolutions from 8-100 Yee cells per cm(3), there is only a 16% variation in maximum SAR in any 1 cm(3) of tissue in the head, and only a 7% variation in SAR averaged over the entire head. CONCLUSION: While it is always desirable to perform SAR calculations with the greatest possible accuracy, in calculations of the maximum SAR levels in any 1 cm(3) of tissue, spatial resolutions greater than 5 mm may not yield notably different results than those with a spatial resolution of 5 mm.
    [Abstract] [Full Text] [Related] [New Search]