These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [Preliminary study on using acousto-optic tunable filter as wavelength selector for atomic absorption spectrometry]. Author: Zhao LW, Zhang YH, Wang MJ, Song DQ, Zhang HQ, Jin QH. Journal: Guang Pu Xue Yu Guang Pu Fen Xi; 2002 Jun; 22(3):472-5. PubMed ID: 12938338. Abstract: An acousto-optic tunable filter (AOTF) is an all-solid-state, electronic monochromator that is based on the diffraction of light by an acoustic wave in an anisotropic crystal. It is a new kind of tunable small filter with narrow band compared with traditional unicolor filter. The filter can diffract incident white light at a specific wavelength when a specific radio frequency is applied into it. An AAS experimental setup using a microwave plasma torch (MPT) as the atomizer, and a visible AOTF as the wavelength selector was developed and the analytical performance was evaluated by determination of Na. The effect on the absorption signal of some operating conditions, including the observation height, the microwave forward power and the carrier and support gas flow rates for MPT, were investigated. The detection limit for Na was shown to be 0.23 microgram.mL-1 and the relative standard deviation was 2.6% (n = 6).[Abstract] [Full Text] [Related] [New Search]