  • Title: Determination of EDTA species in water by square-wave voltammetry using a chitosan-coated glassy carbon electrode.
    Author: Zhao C, Pan Y, Su Y, Zhang Z, Guo Z, Sun L.
    Journal: Water Res; 2003 Oct; 37(17):4270-4. PubMed ID: 12946910.
    Abstract:
    The electrochemical behavior of Fe(EDTA)(-/2-) on a chitosan-coated glassy carbon electrode (GCE) and the determining method of the EDTA species in water samples by square-wave voltammetry (SWV) have been investigated. Fe(EDTA)(-) can be adsorbed on the positively charged chitosan membrane assembled on a GCE and be reduced at a potential of -0.095V vs. Ag/AgCl. On the chitosan-coated GCE, the cathodic stripping peak current is proportional to the concentration of EDTA after the operating parameters of SWV were optimized. The measuring range of EDTA was from 2.0x10(-6) to 8.0x10(-5)mol/L with a correlation coefficient of r=0.998 and a detection limit of 2.0x10(-7)mol/L. The relative standard deviation was less than 6.4% (n=5) and the recovery was in the range of 98-105% for the determination of practical samples. The result was consistent with that from the HPLC method.