  • Title: Seven years of left ventricular pacing due to malposition of pacing electrode.
    Author: Shmuely H, Erdman S, Strasberg B, Rosenfeld JB.
    Journal: Pacing Clin Electrophysiol; 1992 Apr; 15(4 Pt 1):369-72. PubMed ID: 1374879.
    Abstract:
    The case report is presented of a patient in whom an uncomplicated left ventricular transvenous pacing produced right bundle branch block (RBBB). A diagnostic echocardiography, confirmed by cine cardiovascular computed tomography, showed that there was no rupture of the right ventricle and diagnosed a left ventricular pacing, due to malposition of the pacing electrode. The patient was treated with aspirin and dipyridamole during the last 6 years of follow-up, without any complications, including 1 year of pacing, prior to admission.