These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Correction for attenuation in technetium-99m-HMPAO SPECT brain imaging. Author: Kemp BJ, Prato FS, Dean GW, Nicholson RL, Reese L. Journal: J Nucl Med; 1992 Oct; 33(10):1875-80. PubMed ID: 1403161. Abstract: We present a correction technique that uses the effective bone and tissue attenuation coefficients to compensate 99mTc-HMPAO brain SPECT projections for attenuation. Transverse images of a human skull filled with a uniform mixture of 99mTc and gelatin have a greater count density at the center with respect to the periphery when corrected for attenuation with the effective water/tissue coefficient of 0.12 cm-1. An attenuation coefficient of 0.09 cm-1 produces uniform images at the expense of a reduced count density. Additional experiments with phantoms wrapped with aluminum (to simulate bone) indicate that the greater count density at the image center is a result of increased attenuation at the edges of the projections where there is a greater path length through the aluminum (or bone). SPECT projections explicitly corrected for both bone and soft-tissue attenuation result in images of improved uniformity and increased count density.[Abstract] [Full Text] [Related] [New Search]