These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: An experimental apparatus for the simulation of three-dimensional movements in orthodontics.
    Author: Bourauel C, Drescher D, Thier M.
    Journal: J Biomed Eng; 1992 Sep; 14(5):371-8. PubMed ID: 1405553.
    Abstract:
    An apparatus for the study of three-dimensional force systems and the resulting movements during orthodontic treatment is presented. The instrument consists of two force-torque transducers which are capable of recording both forces and torques simultaneously, in all spatial directions. Each sensor has a measuring range of 15N (450 Nmm) and a resolution of 0.02 N (0.5 Nmm) and is mounted on a set of three translational and three rotational stages driven by stepping motors. Positioning accuracy is in the range of 1 micron and 0.01 degrees respectively. The apparatus is computer controlled and supported by comprehensive software.
    [Abstract] [Full Text] [Related] [New Search]