These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Electric field of a six-needle array electrode used in drug and DNA delivery in vivo: analytical versus numerical solution. Author: Dev SB, Dhar D, Krassowska W. Journal: IEEE Trans Biomed Eng; 2003 Nov; 50(11):1296-300. PubMed ID: 14620000. Abstract: We present an analytical solution for the electrical potential and field established by a six-needle array electroporation electrode, which is used in vivo for cancer treatment and DNA delivery. The analytical solution closely matches the numerical solution obtained with the finite element method: the mean error is less than 0.6%.[Abstract] [Full Text] [Related] [New Search]