These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [Particle enhanced immunoturbidimetic assay of high sensitive C-reactive protein]. Author: Hu M, Li YW, Hu YH. Journal: Hunan Yi Ke Da Xue Xue Bao; 2003 Aug; 28(4):415-7. PubMed ID: 14653135. Abstract: OBJECTIVE: To evaluate the particle enhanced immunoturbidimetic assay of high sensitive C-reactive protein by the automated chemistry analyzer (Hitachi 7170A). METHODS: The detection limit of quantification, measurement range, precision, interference and reference interval were assessed. RESULTS: The measurement range of the method was from 0.125 to 12.5 mg/L. The detection limitation was 0.115 mg/L. The internal and external accuracy were 0.89% and 0.90% respectively. The reference range of the assay was from 0.20 mg/L to 2.60 mg/L. CONCLUSION: The particle enhanced immunoturbidimetic assay is sensitive and stable, and can be used in clinical diagnosis.[Abstract] [Full Text] [Related] [New Search]