These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Time-resolved Fourier transform infrared spectroscopic imaging.
    Author: Bhargava R, Levin IW.
    Journal: Appl Spectrosc; 2003 Apr; 57(4):357-66. PubMed ID: 14658631.
    Abstract:
    Fourier transform infrared (FT-IR) imaging allows simultaneous spectral characterization of large spatial areas due to its multichannel detection advantage. The acquisition of large amounts of data in the multichannel configuration results, however, in a poor temporal resolution of sequentially acquired data sets, which limits the examination of dynamic processes to processes that have characteristic time scales of the order of minutes. Here, we introduce the concept and instrumental details of a time-resolved infrared spectroscopic imaging modality that permits the examination of repetitive dynamic processes whose half-lives are of the order of milli-seconds. As an illustration of this implementation of step-scan FT-IR imaging, we examine the molecular responses to external electric-field perturbations of a microscopically heterogeneous polymer-liquid crystal composite. Analysis of the spectroscopic data using conventional univariate and generalized two-dimensional (2D) correlation methods emphasizes an additional capability for accessing of simultaneous spatial and temporal chemical measurements of molecular dynamic processes.
    [Abstract] [Full Text] [Related] [New Search]