These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Real-time determination of picomolar free Cu(II) in seawater using a fluorescence-based fiber optic biosensor. Author: Zeng HH, Thompson RB, Maliwal BP, Fones GR, Moffett JW, Fierke CA. Journal: Anal Chem; 2003 Dec 15; 75(24):6807-12. PubMed ID: 14670039. Abstract: We report real-time, in situ determination of free copper ion at picomolar levels in seawater using a fluorescence-based fiber optic biosensor. The sensor transducer is a protein molecule, site-specifically labeled with a fluorophore that is attached to the distal end of an optical fiber, which binds free Cu(II) with high affinity and selectivity. The transducer reports the metal's concentration as a change in fluorescence intensity or lifetime, using a frequency domain approach. The transducer's response time is diffusion-limited, with a typical measurement requiring 30 s. The sensor demonstrates a detection limit of 0.1 pM free Cu(II) in a seawater model. Accuracy and precision of the sensor were at least comparable to cathodic ligand exchange/adsorptive cathodic stripping voltammetry. Measurements of tidal flushing of a copper-contaminated inlet are shown.[Abstract] [Full Text] [Related] [New Search]