These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Determination of trace amount of carbon disulfide in water by the spectrophotometric reaction-rate method. Author: Ensafi AA, Rahimi Mansour H, Majlesi R. Journal: Anal Sci; 2003 Dec; 19(12):1679-81. PubMed ID: 14696937. Abstract: Carbon disulfide (CS2) was determined at trace levels by its induction effect on the reaction of triiodide with azide in acidic media. The reaction was monitored spectrophotometrically by the decreased absorbance of triiodide over a period of 30 - 180 s from the initiation of the reaction at 350 nm, with calibration linear range of 0.020 - 1.870 microg/ml CS2. The limit of detection is 0.013 microg/ml CS2. A carbon-active column was used for sample solution clean up. The proposed method was successfully applied to the determination of trace carbon disulfide in natural water.[Abstract] [Full Text] [Related] [New Search]