These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Timing of hemodynamic pressure measurements on derived hemodynamic parameters. Author: Urquhart G, Jensen L. Journal: Dynamics; 2003; 14(3):13-20. PubMed ID: 14725142. Abstract: The purpose of this study was to examine the difference between hemodynamic pressures and parameters obtained pre- compared to post-thermodilution CO measurements. A repeated measures within subject design was conducted with a cardiac surgical cohort. Three measures of hemodynamic pressures and parameters were determined pre- and post-CO measurements (Set 1) and repeated in 30 minutes (Set 2). The sequence was duplicated in four hours (Sets 3 and 4). Hemodynamic pressures lower pre-CO were PAS at Sets 1 and 3, and SBP, DBP, and MAP at Set 3. Hemodynamic parameters lower pre-CO were PVRI at Set 1 and SVRI at Set 3. These pre-post CO differences did not vary by greater than 10%. As the CO injectate volume had minimal effect, hemodynamic pressures may be obtained pre- or post-CO to derive hemodynamic parameters.[Abstract] [Full Text] [Related] [New Search]