These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: In-line monitoring of esterification using a miniaturised mid-infrared spectrometer.
    Author: McGill CA, Ferguson RH, Donoghue K, Nordon A, Littlejohn D.
    Journal: Analyst; 2003 Dec; 128(12):1467-70. PubMed ID: 14737234.
    Abstract:
    A miniaturised mid-infrared (MIR) spectrometer attached directly to an in-line immersion probe with an attenuated total reflectance crystal has been used to monitor the progression of the batch reaction of crotonic acid and 2-butanol in toluene. Univariate calibration, using the signal at 1188 cm(-1) in the 2nd derivative spectrum, gave accurate (average error, 1.6%) and precise (average relative standard deviation, 5.2%) estimation of 2-butyl crotonate concentrations in the range 0.08-0.49 mol dm(-3). Calibration by partial least squares was of no additional benefit in this application. The performance of in-line MIR spectrometry was comparable to that of an off-line reference gas chromatography method and superior to that of other in-line process analysis techniques (near-infrared, Raman or UV-visible spectrometries).
    [Abstract] [Full Text] [Related] [New Search]