These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Accuracy of noncycloplegic autorefraction in school-age children in China. Author: Zhao J, Mao J, Luo R, Li F, Pokharel GP, Ellwein LB. Journal: Optom Vis Sci; 2004 Jan; 81(1):49-55. PubMed ID: 14747761. Abstract: PURPOSE: To evaluate the accuracy of noncycloplegic autorefraction in a representative sample of school-age children in China. METHODS: Refractive error was measured with an autorefractor, both before and after cycloplegia induced with cyclopentolate, in a population-based sample of 4973 children between the ages of 7 and 18 years. Spherical equivalent refractive error and astigmatism as represented by Jackson crossed-cylinders (J0 and J45) were the main outcome measures. RESULTS: Noncycloplegic measurements of equivalent spheres were consistently more negative or less positive than those after cycloplegia, with mean +/- SD differences of -1.23 +/- 0.97 D. The differences were particularly large for hyperopic eyes (mean difference of -2.98 +/- 1.65 D for hyperopia of at least +2.00 D) while becoming progressively smaller for emmetropic eyes, and smaller yet for myopic eyes (mean difference of -0.41 +/- 0.46 D for myopia of -2.00 D or more). Increasing age was associated with increased, but clinically insignificant, differences. Little difference was found between noncycloplegic and cycloplegic measurements of astigmatism: mean J0 and J45 differences were -0.08 +/- 0.13 D and -0.01 +/- 0.09 D, respectively. CONCLUSIONS: Noncycloplegic autorefraction was found to be highly inaccurate in school-age children and, thus, not suitable for studies of refractive error or for prescription of glasses in this population.[Abstract] [Full Text] [Related] [New Search]