These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Percutaneous retrieval of chronic intravascular foreign bodies. Author: Savage C, Ozkan OS, Walser EM, Wang D, Zwischenberger JB. Journal: Cardiovasc Intervent Radiol; 2003; 26(5):440-2. PubMed ID: 14753301. Abstract: To evaluate the feasibility of intravascular retrieval of chronic foreign bodies, we retrospectively reviewed an 8 year experience (1993-2001) of percutaneous retrieval of chronically retained intravascular foreign bodies (n = 6). In 6 of 6 cases (4 catheter fragments, 2 guidewires), 5-90 days elapsed before retrieval via the femoral or internal jugular vein. Under fluoroscopy, we determined the foreign body's course, position and size. A guidewire was advanced through a multipurpose catheter to the foreign body. The multipurpose catheter was replaced with a gooseneck snare catheter and the snare advanced to grasp and remove the foreign body. Percutaneous retrieval was successful in all 6 cases. One patient experienced mild hemoptysis, which resolved within 24 hr of observation. No patient experienced long-term sequelae. Given the potential life-threatening complications from intravascular foreign bodies and the low complication rate from percutaneous retrieval, we recommend extraction of the foreign body even if it is asymptomatic in the chronic setting (> 24 hr).[Abstract] [Full Text] [Related] [New Search]