These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [Thermoluminescent dosimetry for radiation exposure of the breast during film and xeromammography (author's transl)]. Author: Säbel M, Ruff A, Weishaar J. Journal: Rofo; 1978 May; 128(5):616-22. PubMed ID: 149076. Abstract: Measurements of radiation exposure of the breast during various mammographic procedures were carried out with LiF dosimeters. Using non-screen film (Kodak Definix Medical) at 30 kv., average skin dose in 105 patients was 10.1 R per exposure. During xeroradiography (45 kv.) skin dose was about 25% of this and for a film-screen system (Mammoray RP 3/MR 50 at 30 kv.) only 7% of this value. Skin dose (at constant potential) rises almost exponentially with increasing tissue thickness. The age of the patient and the type of tissue in the breast has a significant effect on the radiation dose. From the measurements, an average dose per exposure was calculated for the entire breast and for the glandular tissue. This gave a result of 1.6 rd (16 mGy) and 1.0 rd (10 mGy).[Abstract] [Full Text] [Related] [New Search]