These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Aqueous sample in an EPR cavity: sensitivity considerations.
    Author: Nesmelov YE, Gopinath A, Thomas DD.
    Journal: J Magn Reson; 2004 Mar; 167(1):138-46. PubMed ID: 14987608.
    Abstract:
    The radial mode matching (RMM) method has been used to calculate accurately the microwave field distribution of the TE(011) mode in a spherical EPR cavity containing a linear aqueous sample, in order to understand in detail the factors affecting sensitivity in EPR measurements at X band. Specific details of the experiment were included in the calculations, such as the cavity geometry, the presence of a quartz dewar, the size of the aqueous sample, and the sample's dielectric properties. From the field distribution, several key physical parameters were calculated, including cavity Q, filling factor, mean microwave magnetic field at the sample, and cavity efficiency parameter Lambda. The dependence of EPR signal intensity on sample diameter for a cylindrical aqueous sample was calculated and measured experimentally for non-saturated and half-saturated samples. The optimal aqueous sample diameter was determined for both cases. The impact of sample temperature, conductivity, and cavity Q on sensitivity of EPR is discussed.
    [Abstract] [Full Text] [Related] [New Search]