These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Highly precise measurements of the relative gamma-ray intensities for 56Mn and 72Ga. Author: Miyahara H, Ogata Y, Fujiki K, Katoh K, Marnada N. Journal: Appl Radiat Isot; 2004; 60(2-4):295-9. PubMed ID: 14987657. Abstract: Precise relative gamma-ray intensities of 56Mn and 72Ga produced by neutron irradiation of natural Mn3O4 and enriched 71Ga2O3 (99.8%) were measured by an HPGe detector with highly-precisely calibrated detection efficiencies. As both nuclides exhibit no beta branch to the ground state, the absolute emission probabilities of the 846.8 keV gamma ray for 56Mn and 834.0 keV gamma ray for 72Ga were determined to be 0.9888+/-0.0014 and 0.9553+/-0.0005, respectively.[Abstract] [Full Text] [Related] [New Search]