These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Evaluation of the new EMIT tacrolimus assay in kidney and liver transplant recipients.
    Author: Akbas SH, Yavuz A, Tuncer M, Yurdakonar E, Akcit F, Gurkan A, Demirbas A, Gultekin M, Ersoy F, Akaydin M.
    Journal: Transplant Proc; 2004; 36(1):86-8. PubMed ID: 15013308.
    Abstract:
    Tacrolimus (FK506) is a potent macrolide immunosuppressant used for prevention of organ transplant rejection following transplantation. Monitoring of blood tacrolimus concentrations is essential to assess organ rejection and toxicity, because of the agent's narrow therapeutic range, wide inter- and intraindividual pharmacokinetic variability as well as drug interactions mediated by alteration in cytochrome P450. Several methods have been developed to monitor tacrolimus; immunoassays, bioassays, and HPLC/MS. The purpose of this study was to compare two analytical methods: the well-established MEIA II tacrolimus immunoassay using the IMx analyzer and the new EMIT 2000 tacrolimus immunoassay on the Cobas Integra 400 system. Tacrolimus results obtained using the two methods have been compared on 180 whole blood samples from kidney and liver transplant patients. The analytical sensitivities of both methods were defined as 1.2 ng/mL for EMIT and 1.5 ng/mL for MEIA II. The within-run CVs (n = 15) obtained with four-level controls were 9.08%, 9.41%, 5.23% and 4.4% for EMIT 2000. The comparison showed the following relationship between two methods: MEIA = 1.08.EMIT + 0.20 (r =.893). In conclusion, the EMIT 2000 tacrolimus immunoassay is a reliable alternative for the MEIA II method to monitor tacrolimus in organ transplant recipients. It provides a valid quantitative measurement of tacrolimus with comparable % CVs in quality-control as well as patient blood samples. Additionally, the EMIT 2000 method provides a rapid analysis of a large number of samples in one run with a low turnaround time and possibilities to reanalyze critical samples.
    [Abstract] [Full Text] [Related] [New Search]