These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Determination of nimodipine in human plasma by a sensitive and selective liquid chromatography-tandem mass spectrometry method. Author: Qiu F, Chen X, Li X, Zhong D. Journal: J Chromatogr B Analyt Technol Biomed Life Sci; 2004 Apr 05; 802(2):291-7. PubMed ID: 15018790. Abstract: A sensitive and highly selective liquid chromatography-tandem mass spectrometry (LC-MS-MS) method was developed to determine nimodipine in human plasma. The analyte and internal standard nitrendipine were extracted from plasma samples by n-hexane-dichloromethane-isopropanol (300:150:4, v/v/v), and chromatographed on a C(18) column. The mobile phase consisted of methanol-water-formic acid (80:20:1, v/v/v). Detection was performed on a triple quadrupole tandem mass spectrometer by selected reaction monitoring (SRM) mode via atmospheric pressure chemical ionization (APCI) source. The method has a limit of quantification of 0.24 ng/ml. The linear calibration curves were obtained in the concentration range of 0.24-80 ng/ml. The intra- and inter-day precisions were lower than 4.4% in terms of relative standard deviation (R.S.D.), and the accuracy ranged from 0.0 to 5.8% in terms of relative error (RE). This validated method was successfully applied for the evaluation of pharmacokinetic profiles of nimodipine tablets administered to 18 healthy volunteers.[Abstract] [Full Text] [Related] [New Search]