These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Canine and feline hematology reference values for the ADVIA 120 hematology system.
    Author: Moritz A, Fickenscher Y, Meyer K, Failing K, Weiss DJ.
    Journal: Vet Clin Pathol; 2004; 33(1):32-8. PubMed ID: 15048625.
    Abstract:
    BACKGROUND: The ADVIA 120 is a laser-based hematology analyzer with software applications for animal species. Accurate reference values would be useful for the assessment of new hematologic parameters and for interlaboratory comparisons. OBJECTIVE: The goal of this study was to establish reference intervals for CBC results and new parameters for RBC morphology, reticulocytes, and platelets in healthy dogs and cats using the ADVIA 120 hematology system. METHODS: The ADVIA 120, with multispecies software (version 1.107-MS), was used to analyze whole blood samples from clinically healthy dogs (n=46) and cats (n=61). Data distribution was determined and reference intervals were calculated as 2.5 to 97.5 percentiles and 25 to 75 percentiles. RESULTS: Most data showed Gaussian or log-normal distribution. The numbers of RBCs falling outside the normocytic-normochromic range were slightly higher in cats than in dogs. Both dogs and cats had reticulocytes with low, medium, and high absorbance. Mean numbers of large platelets and platelet clumps were higher in cats compared with dogs. CONCLUSIONS: Reference intervals obtained on the ADVIA 120 provide valuable baseline information for assessing new hematologic parameters and for interlaboratory comparisons. Differences compared with previously published reference values can be attributed largely to differences in methodology.
    [Abstract] [Full Text] [Related] [New Search]