These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Therapeutic drug monitoring for sirolimus in whole blood of organ transplants by high-performance liquid chromatography with ultraviolet detection. Author: Campanero MA, Cardenas E, Sádaba B, García-Quetglas E, Muñoz-Juarez MJ, Gil-Aldea I, Pazo D, Azanza JR, Honorato J. Journal: J Chromatogr A; 2004 Mar 26; 1031(1-2):265-73. PubMed ID: 15058592. Abstract: We developed and validated an accurate, sensitive, precise and rapid HPLC method with UV detection for the determination of sirolimus in blood samples from renal, cardiac and hepatic transplants. This method overcomes most of the problems related to previously published assays using a narrow-bore column with base deactivated C18 reversed phase. Whole blood samples were purified by a combination of a precipitating blood matrix with zinc sulphate and a single step liquid-liquid extraction with acetone and 1-chlorobutane. Calibration curves (range 2.5-150 ng/ml), were linear with coefficients of correlation better than 0.996. The relative standard deviation was determined to be less than 8%. The present method has also been validated by a reference laboratory (St. George's Hospital Medical School, London, UK). More of 300 clinical samples have been analysed with this method.[Abstract] [Full Text] [Related] [New Search]