  • Title: [Clinical evaluation of stent placement for tracheal and bronchial stenosis].
    Author: Fujihara Y, Sawada S, Tanabe Y, Koyama T, Iwamiya K, Yoshida K, Kato T, Irizawa T, Nakai I, Furui S.
    Journal: Nihon Igaku Hoshasen Gakkai Zasshi; 1992 May 25; 52(5):576-88. PubMed ID: 1508631.
    Abstract:
    Expandable metallic stents were successfully introduced in 7 patients, including 4 with left main bronchial stenosis caused by bronchopulmonary tuberculosis, 2 with main bronchial stenosis caused by lung cancer and one with tracheal stenosis caused by adenoid cystic carcinoma. The length of stenosis was 1.5-5 cm. The stents were 1.5-2.5 cm long with barbs, and their full expanded diameter was 1.5 cm. Balloon dilatation was performed before stenting in all cases. The stents were inserted by using a 10-12 Fr catheter. In all patients except the one with tracheal stenosis, stents were introduced under local anesthesia without any difficulties. No migration of stents occurred. After stent placement, there were no respiratory difficulties, and radionuclide lung perfusion scan and chest radiographic findings such as lung atelectasis showed marked improvement in three cases. Combined therapy of stent placement and bronchial arterial infusion chemotherapy showed marked effectiveness in one case with lung cancer. Expandable metallic stents were very useful in eliminating tracheobronchial stenosis symptoms.