These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: A comparison of mammography screen-film combinations. Author: Schueler BA, Gray JE, Gisvold JJ. Journal: Radiology; 1992 Sep; 184(3):629-34. PubMed ID: 1509043. Abstract: A comparison study was performed to evaluate the image quality and radiation dose of six mammographic screen-film combinations: a Min-R screen with OM-1, SO-155, and SO-177 films (Eastman Kodak); a Min-R medium screen (Eastman Kodak) with OM-1 film; an HR Mammo medium screen (Fuji Medical Systems USA) with OM-1 film; and a Min-R fast screen with T-Mat M II film (Eastman Kodak). SO-177 films were processed with an extended cycle. Exposures of an acrylic test object with embedded masses, fibers, and specks and of a preserved breast specimen were made, for two paired image comparison tests in which the visibility of diagnostic features, contrast, and noise were judged. In most areas of image quality evaluated, a Min-R screen with OM-1, SO-155, and SO-177 films was superior. These three screen-film combinations had similar imaging characteristics, even though OM-1 film requires a higher radiation exposure. Images produced with a Min-R fast screen and T-Mat M II film were significantly lower in quality.[Abstract] [Full Text] [Related] [New Search]