These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: NEMA NU 2 performance tests for scanners with intrinsic radioactivity. Author: Watson CC, Casey ME, Eriksson L, Mulnix T, Adams D, Bendriem B. Journal: J Nucl Med; 2004 May; 45(5):822-6. PubMed ID: 15136632. Abstract: Performance tests on lutetium oxyorthosilicate (LSO)-based PET scanners cannot be conducted strictly according to the National Electrical Manufacturers Association (NEMA) NU 2 standards because of the presence of intrinsic radioactivity within the LSO crystal scintillator material. This background radiation gives rise mainly to random coincidence events but also to a small number of true coincidences, which cannot be eliminated from measurements on such scanners and must therefore be corrected for in the data analysis. The current NU 2 standards do not take account of these backgrounds and hence can lead to erroneous results on LSO-based machines. Nevertheless, the intent of the standards can be met with appropriate modifications to the acquisition and processing procedures. In this paper, we propose certain changes to the NEMA specifications to accommodate this class of scanners. These changes affect mainly the estimation of sensitivity, scatter, randoms, and count losses. Using these modified procedures, the NU 2 performance of LSO-based systems can accurately be measured.[Abstract] [Full Text] [Related] [New Search]