These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Determination of beta-lactams in milk using a surface plasmon resonance-based biosensor. Author: Gustavsson E, Degelaen J, Bjurling P, Sternesjö A. Journal: J Agric Food Chem; 2004 May 19; 52(10):2791-6. PubMed ID: 15137815. Abstract: Two surface plasmon resonance (SPR)-based biosensor assays for detection of beta-lactam antibiotics in milk are reported. The assays are based on the enzymatic activity of a carboxypeptidase converting a 3-peptide into a 2-peptide, a reaction that is inhibited in the presence of beta-lactams. Antibodies were used to measure either the amount of formed enzymatic product or the amount of remaining enzymatic substrate. Both assays detected different beta-lactams at or below European maximum residue limits (MRLs), and the detection limit for penicillin G was 1.2 microg/kg and 1.5 microg/kg for the 2- and 3-peptide assays, respectively. The precision (CV) was < 5%, both within and between assays at the penicillin G MRL (4 microg/kg). The biosensor results obtained upon analysis of incurred milk samples were compared with results obtained by liquid chromatography (HPLC), and the method agreements were, in general, good.[Abstract] [Full Text] [Related] [New Search]