These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Thin film trichloroethylene electrochemical sensor. Author: Chen MH, Liu CC, Chou TC. Journal: Biosens Bioelectron; 2004 Jul 30; 20(1):25-32. PubMed ID: 15142573. Abstract: Pt-Ti and Pb-Pt-Ti thin films were deposited on alumina substrates by sputtering in Ar gas. In this study, an electrodeposited Pb-modified Pt-Ti thin film working electrode was prepared. Optimal sensing conditions were found to be -2.10 V (versus Ag/Ag+ with 0.1 M tetrabutylammonium perchlorate (TBAP) in acetonitrile (AN) solution) sensing potential, 250 rpm agitation rate. At room temperature, the response time was 15 s (90% response time). The correlation of sensing response current, id, and trichloroethylene (TCE) concentration, CL, is id = 2.86CL in the range from 100 to 700 ppm TCE. Additionally, the rate constant of (TCE) cathodic reduction was found to be 2.434 x 10(-3) cm(-1) s(-1).[Abstract] [Full Text] [Related] [New Search]