  • Title: Thin film trichloroethylene electrochemical sensor.
    Author: Chen MH, Liu CC, Chou TC.
    Journal: Biosens Bioelectron; 2004 Jul 30; 20(1):25-32. PubMed ID: 15142573.
    Abstract:
    Pt-Ti and Pb-Pt-Ti thin films were deposited on alumina substrates by sputtering in Ar gas. In this study, an electrodeposited Pb-modified Pt-Ti thin film working electrode was prepared. Optimal sensing conditions were found to be -2.10 V (versus Ag/Ag+ with 0.1 M tetrabutylammonium perchlorate (TBAP) in acetonitrile (AN) solution) sensing potential, 250 rpm agitation rate. At room temperature, the response time was 15 s (90% response time). The correlation of sensing response current, id, and trichloroethylene (TCE) concentration, CL, is id = 2.86CL in the range from 100 to 700 ppm TCE. Additionally, the rate constant of (TCE) cathodic reduction was found to be 2.434 x 10(-3) cm(-1) s(-1).