  • Title: Truncation correction of fan beam transmission data for attenuation correction using parallel beam emission data on a 3-detector SPECT system.
    Author: Takahashi Y, Murase K, Mochizuki T, Higashino H, Motomura N.
    Journal: Nucl Med Commun; 2004 Jun; 25(6):623-30. PubMed ID: 15167524.
    Abstract:
    BACKGROUND: When the simultaneous transmission computed tomography (TCT)/single photon emission CT (SPECT) acquisition protocol is applied to myocardial studies using a 3-detector SPECT, the narrow effective field of view of a fan beam collimator used for TCT acquisition may cause truncation artifacts on TCT images. In this paper, we propose a new method of correcting for the truncation of TCT. METHODS: The truncated parts of the TCT projection data are corrected using quadratic functions, based on the properties that the integral of non-truncated TCT projection data is constant at any projection angle and the position of the centre of gravity is focused on a fixed point. The usefulness of our method was investigated in phantom and human studies using a 3-detector SPECT equipped with one cardiac fan beam collimator for TCT and two parallel beam collimators for SPECT. We used Tl as a tracer for SPECT and Tc as an external source for TCT. RESULTS: The phantom and human studies showed that our method can adequately correct for the truncation of TCT data acquired using a fan beam collimator in a 3-detector SPECT, as long as there is no truncation in SPECT data. CONCLUSION: Our method appears to be useful for improving the SPECT images obtained using simultaneous TCT/SPECT acquisition in a 3-detector SPECT. However, further studies will be necessary to establish the clinical usefulness of this method.
    [Abstract] [Full Text] [Related] [New Search]