These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Conflicting results between electrophoresis methods of serum M-proteins. Author: Wuyts B, Bossuyt X, Verhoef G, Blanckaert N, Delanghe JR. Journal: Electrophoresis; 2004 Jun; 25(10-11):1548-50. PubMed ID: 15188240. Abstract: The quantification of monoclonal immunoglobulins by protein electrophoresis has been helpful in determining the prognosis for the patient. In a 65-year old man with documented IgG lambda monoclonal gammopathy, a discrepancy in the detection and quantification of this M-protein was found when studied by four different electrophoretic methods. On a high-resolution agarose method with acid violet staining a prominent peak was seen in the fast gamma-region. A lower resolution agarose method using an amido black staining showed a small mid-gamma restriction, and no spike was detected by high-resolution or lower resolution capillary zone electrophoresis. The discrepancy was not attributable to a migration in the beta-region leading to masking by transferrin or C3 peak, incorrect separation on CZE due to a high isoelectric point (pI > 8.5) or pentamerization of IgM M-proteins. Therefore, other causes for discrepancies should be investigated.[Abstract] [Full Text] [Related] [New Search]