These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Evaluation of an adaptive directional system in a DSP hearing aid.
    Author: Bentler RA, Tubbs JL, Egge JL, Flamme GA, Dittberner AB.
    Journal: Am J Audiol; 2004 Jun; 13(1):73-9. PubMed ID: 15248806.
    Abstract:
    The effectiveness of an adaptive directional microphone design, as implemented in the Phonak Claro behind-the-ear hearing aid, is evaluated. Participants were fit bilaterally and tested in 2 environments, an anechoic chamber and a moderately reverberant classroom, with the microphones in the fixed (cardioid) setting and the adaptive setting. Five speakers were placed between 110 degrees and 250 degrees azimuth around the listener. Speech-weighted noise was presented from those speakers at an overall level (OAL) of 65 dB (A). Noise was increased by 8 dB from 1 speaker at a time, using 2-s modulation and random assignment, while the output from the other speakers was reduced to maintain the constant OAL. Results of 2 speech perception tasks used as outcome measures indicated that the adaptive system was not able to follow the dominant noise source in the presence of lower level noise sources. Self-report measures obtained after blinded home trials were consistent with laboratory findings that the participants did not perceive this adaptive microphone design to be more effective than the default fixed-microphone option.
    [Abstract] [Full Text] [Related] [New Search]