These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Electrochemical detection of ascorbic acid (vitamin C) using a glassy carbon electrode.
    Author: Ly SY, Chae JI, Jung YS, Jung WW, Lee HJ, Lee SH.
    Journal: Nahrung; 2004 Jun; 48(3):201-4. PubMed ID: 15285111.
    Abstract:
    A method for the detection of ascorbic acid (AA) was developed using square wave anodic stripping voltammetric (SWASV) analysis with a glassy carbon electrode (GCE). The results indicated a sensitive oxidation peak current of AA on the GCE. A linear curve was obtained within a concentration range of 1-130.0 microg/L with a preconcentration time of 270 s. The relative standard deviation of 30.0 microg/L observed was 0.1016% (n = 12) under optimum conditions. The low detection limit (S/N) was pegged at 0.30 microg/L. Results showed that the method developed can be used to assay biological and pharmaceutical samples, and food samples, as well as other materials requiring AA analyses.
    [Abstract] [Full Text] [Related] [New Search]