These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Optical coherence tomography system with no high-precision scanning stage and stage controller. Author: Lo YL, Kuo CI, Chuang CH, Yan ZZ. Journal: Appl Opt; 2004 Jul 20; 43(21):4142-9. PubMed ID: 15291056. Abstract: Optical coherence tomography (OCT) is a novel technique for noninvasive imaging based on the use of a low-coherence interferometer. Conventionally, obtaining high-resolution images requires the use of high-precision sample and scanning stages and a stage controller for simultaneous measurement of the refractive index and the thickness of an optical sample. However, in this study a novel optical-fiber-type OCT system is developed that does not need both a high-precision scanning stage and a stage controller. Additionally, two signal demodulation processes are described. Compared with that of conventional OCT systems, the current configuration eliminates the high-precision scanning stage and stage controller and is therefore cheaper and less complex. Also, this new technique could be applied to conventional OCTs in biotissue scanning.[Abstract] [Full Text] [Related] [New Search]