These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Efficacy of lower-energy biphasic shocks for transthoracic defibrillation: a follow-up clinical study. Author: Higgins SL, O'Grady SG, Banville I, Chapman FW, Schmitt PW, Lank P, Walker RG, Ilina M. Journal: Prehosp Emerg Care; 2004; 8(3):262-7. PubMed ID: 15295725. Abstract: OBJECTIVE: This clinical study prospectively evaluated the first-shock defibrillation efficacy of 150-joule impedance-compensated, 200-microF biphasic truncated exponential (BTE) shocks in patients with electrically-induced ventricular fibrillation (VF), and compared it with a historical control group treated with 200-J monophasic damped sine (MDS) shocks. METHODS: Ventricular tachyarrhythmias were induced in patients undergoing electrophysiologic (EP) testing for ventricular arrhythmias or testing of an implantable cardioverter-defibrillator (ICD). A 150-J shock was delivered as the primary therapy to terminate induced arrhythmias in the EP group, and as a "rescue" shock when a single ICD shock failed to terminate the arrhythmias in the ICD group. RESULTS: Ninety-six patients received study shocks. The preshock rhythm was classified as VF in 77 patients and as ventricular tachycardia (VT) in 19 patients. First-shock success rates for VF and VT were 75 out of 77 (97.4%) and 19 out of 19 (100%) for the 150-J BTE compared with the historical control rates of 61 out of 68 (89.7%) and 29 out of 31 (94%) for 200-J MDS. The first-shock success rate for VF treated with 150-J BTE was technically equivalent to that of 200-J MDS (p=0.001). The transthoracic impedance did not vary between groups, yet the peak current delivered by the 150-J BTE shock was about 50% lower. CONCLUSIONS: This study demonstrated that 150-J shocks of this impedance-compensated, 200-microF BTE waveform provided very high efficacy for defibrillation of short duration, electrically-induced VF. These lower-energy biphasic shocks had a success rate equivalent to that of 200-J MDS shocks, and they provided this efficacy while exposing patients to much less current than the monophasic shocks.[Abstract] [Full Text] [Related] [New Search]