These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Stratus CS cardiac troponin I method: performance characteristics including imprecision at low concentrations.
    Author: Christenson RH, Cervelli DR, Bauer RS, Gordon M.
    Journal: Clin Biochem; 2004 Aug; 37(8):679-83. PubMed ID: 15302610.
    Abstract:
    OBJECTIVES: Define performance characteristics of the Stratus CS cardiac troponin I (cTnI) assay in routine practice. METHODS: Imprecision at low cTnI values was assessed according to NCCLS EP5-A. Linearity was assessed by regression analysis; the reference interval was determined in 345 healthy individuals. Minimum detectable concentration (MDC) was assessed with 20 replicates of zero calibrator, and method correlation and bias (n = 77) were by Passing-Bablok regression and Bland-Altman plot. RESULTS: cTnI of 0.06 microg/l corresponded to the 10% CV; the 99th percentile of the reference population was 0.07 microg/l. Linearity extended from 0 to 50.0 microg/l and the MDC was 0.010 microg/l. The method comparison equation was Stratus CS = 1.00 (Dimension RxL)-0.02 microg/l with no bias. CONCLUSIONS: The Stratus CS cTnI assay in routine practice has performance characteristics appropriate for clinical use, including good correlation to a central laboratory cTnI method and imprecision of a high sensitivity troponin method with a CV of <10% at the 99th percentile of the reference population.
    [Abstract] [Full Text] [Related] [New Search]