These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: K-edge digital subtraction imaging based on a dichromatic and compact x-ray source.
    Author: Sarnelli A, Taibi A, Tuffanelli A, Baldazzi G, Bollini D, Cabal Rodriguez AE, Gombia M, Prino F, Ramello L, Tomassi E, Gambaccini M.
    Journal: Phys Med Biol; 2004 Jul 21; 49(14):3291-305. PubMed ID: 15357198.
    Abstract:
    This work proposes a compact dichromatic imaging system for the application of the K-edge digital subtraction technique based on a conventional x-ray tube and a monochromator system. A quasi-monochromatic x-ray beam at the energy of iodine K-edge is produced by Bragg diffraction on a mosaic crystal. Two thin adjacent beams with energies that bracket the K-edge discontinuity are obtained from the diffracted beam by means of a proper collimation system. They are then detected using an array of Si detectors. A home-made phantom is used to study the image quality as a function of iodine concentration. Signal and signal-to-noise ratio analysis has also been performed. The results are compared with theoretical expectations.
    [Abstract] [Full Text] [Related] [New Search]