These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Adapting the Lagrangian speckle model estimator for endovascular elastography: theory and validation with simulated radio-frequency data.
    Author: Maurice RL, Ohayon J, Finet G, Cloutier G.
    Journal: J Acoust Soc Am; 2004 Aug; 116(2):1276-86. PubMed ID: 15376693.
    Abstract:
    Intravascular ultrasound (IVUS) is known to be the reference tool for preoperative vessel lesion assessments and for endovascular therapy planning. Nevertheless, IVUS echograms only provide subjective information about vessel wall lesions. Since changes in the vascular tissue stiffness are characteristic of vessel pathologies, catheter-based endovascular ultrasound elastography (EVE) has been proposed in the literature as a method for outlining the elastic properties of vessel walls. In this paper, the Lagrangian Speckle Model Estimator (LSME) is formulated for investigations in EVE, i.e., using a polar coordinate system. The method was implemented through an adapted version of the Levenberg-Marquardt minimization algorithm, using the optical flow equations to compute the Jacobbian matrix. The theoretical framework was validated with simulated ultrasound rf data of mechanically complex vessel wall pathologies. The results, corroborated with Ansys finite element software, demonstrated the potential of EVE to provide useful information about the heterogeneous nature of atherosclerotic plaques.
    [Abstract] [Full Text] [Related] [New Search]