These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Inaccuracies in MTS assays: major distorting effects of medium, serum albumin, and fatty acids. Author: Huang KT, Chen YH, Walker AM. Journal: Biotechniques; 2004 Sep; 37(3):406, 408, 410-2. PubMed ID: 15470895. Abstract: Soluble formazan assays are widely used for cell number assessment. However, in our hands, we observed frequent occasions in which the actual cell number was at odds with the assay reading. In this study, we have determined that (i) a large proportion of the reading obtained in commonly used culture media can be caused by media component amplification of formazan production in a way that cannot be corrected for by media-only controls; (ii) the albumin present in 10% serum can reduce the assay absorbance by 50% so that an actual doubling of cell number can be obscured; and (iii) this latter effect is dependent on the concentration of fatty acids. To counter these problems, we have developed a protocol that gives consistent readings that are fully representative of cell number while retaining some of the original advantages of soluble formazan assays.[Abstract] [Full Text] [Related] [New Search]