  • Title: Axial strain calculation using a low-pass digital differentiator in ultrasound elastography.
    Author: Luo J, Bai J, He P, Ying K.
    Journal: IEEE Trans Ultrason Ferroelectr Freq Control; 2004 Sep; 51(9):1119-27. PubMed ID: 15478973.
    Abstract:
    In ultrasound elastography, tissue axial strains are calculated from the gradient of the estimated axial displacements. However, the common differentiation operation amplifies the noises in the displacement estimation, especially at high frequencies. In this paper, a low-pass digital differentiator (LPDD) is proposed to calculate the axial strain from the estimated tissue displacement. Several LPDDs that have been well developed in the field of digital signal processing are presented. The corresponding performances are compared qualitatively and quantitatively in computer simulations and in preliminary phantom and in vitro experiments. The results are consistent with the theoretical analysis of the LPDDs.