These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Determination of optimum planning target volume margins for various tumor sites using electronic portal imaging.
    Author: Andrieu MN, Dirican B, Ozturk AY, Hicsonmez A, Kurtman C.
    Journal: Exp Oncol; 2004 Sep; 26(3):236-9. PubMed ID: 15494694.
    Abstract:
    OBJECTIVE: The aim of this study is to evaluate the efficacy of electronic portal imaging (EPI) to measure the set-up errors for four different sites of irradiation caused by patient positioning. METHODS: A total number of 95 portal images of 11 patients (3 pelvic, 1 total cranium, 3 mantle and 4 tangential fields for breast) were collected during the course of study. The first portal images after a correction of set-up errors according to the simulation films were accepted as the reference images for the subsequent sessions. By matching each portal image with the reference image, the deviations in lateral (x) and superior-inferior (y) axis for all and additionally in antero-posterior (z) axis for pelvis, and standard deviations were calculated. RESULTS: The set-up errors caused by patient's positioning are completely abolished in 15 mm planning target volume (PTV) margins for all studied cases. CONCLUSION: Standard PTV margins usually completely cover the set-up errors caused by patient's positioning.
    [Abstract] [Full Text] [Related] [New Search]