These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: A rapid, sensitive and validated method for the determination of ondansetron in human plasma by reversed-phase high-pressure liquid chromatography. Author: Chandrasekar D, Ramakrishna S, Diwan PV. Journal: Arzneimittelforschung; 2004; 54(10):655-9. PubMed ID: 15553104. Abstract: A simple and sensitive method for the determination of ondansetron (CAS 116002-70-1) in human plasma was developed using high-pressure liquid chromatography (HPLC). The procedure involves extraction of human plasma with tertiary butyl methyl ether containing 2 mol/l sodium hydroxide, followed by reversed-phase HPLC using a LiChrospher 100 RP-18e 5 microm column and UV detection at 305 nm. The retention times of ondansetron and internal standard (propranolol hydrochloride, CAS 318-98-9) were 9.38 and 13.40 min, respectively. The calibration curves were linear over the range of 10 ng/ml (lower limit of quantitation, LOQ) and 380 ng/ml for ondansetron. The intra- and inter-assay coefficients of variation for all the criteria of validation were less than 15% over the linearity range. Ondansetron was stable upon storage in human plasma. The sensitivity and precision of the method were within the accepted limits (< 15 %) throughout the validation period. The present method is useful for determination of plasma concentrations of ondansetron during human pharmacokinetic studies.[Abstract] [Full Text] [Related] [New Search]