These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Comparison of intraocular pressure measurements with the digital tonometer TGDc-01 'PRA' and the Goldmann applanation tonometer.
    Author: Meyer MW, Gockeln R, Hoy L, Meyer A, Erb C.
    Journal: Ophthalmic Res; 2004; 36(5):250-4. PubMed ID: 15583430.
    Abstract:
    BACKGROUND: To compare intraocular pressure (IOP) measurements obtained with the digital tonometer TGDc-01 'PRA' with those from a Goldmann applanation tonometer (GAT). METHODS: The IOP in 176 eyes of 88 healthy volunteers was measured prospectively in a sitting position. One single measurement, generated by the TGDc-01 PRA, was compared with a single reading from the GAT. RESULTS: Mean IOP values were 13.0 +/- 3.7 mm Hg for the TGDc-01 PRA (range, 4-22 mm Hg) and 14.9 +/- 3.2 mm Hg for the GAT (range, 8-27 mm Hg). The mean difference was 1.9 mm Hg with a standard deviation of 2.77 mm Hg, and this was statistically significant (p < 0.001, paired t test). CONCLUSIONS: In comparison to the GAT, the TGDc-01 PRA underestimated IOP on an average of 1.9 mm Hg. Only 50.6% of all measurements were within the +/-2 mm Hg difference range. Thus, the TGDc-01 PRA has no high coincidence degree with the GAT. Both methods were not equivalent.
    [Abstract] [Full Text] [Related] [New Search]