These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Development of high quantum efficiency, flat panel, thick detectors for megavoltage x-ray imaging: a novel direct-conversion design and its feasibility.
    Author: Pang G, Rowlands JA.
    Journal: Med Phys; 2004 Nov; 31(11):3004-16. PubMed ID: 15587653.
    Abstract:
    Most electronic portal imaging devices (EPIDs) developed to date, including recently developed flat panel systems, have low x-ray absorption, i.e., low quantum efficiency (QE) of 2%-4% as compared to the theoretical limit of 100%. A significant increase of QE is desirable for applications such as a megavoltage cone-beam computed tomography (MVCT) and megavoltage fluoroscopy. However, the spatial resolution of an imaging system usually decreases significantly with an increase of QE. The key to the success in the design of a high QE detector is therefore to maintain the spatial resolution. Recently, we demonstrated theoretically that it is possible to design a portal imaging detector with both high QE and high resolution [see Pang and Rowlands, Med. Phys. 29, 2274 (2002)]. In this paper, we introduce such a novel design consisting of a large number of microstructured plates (made by, e.g., photolithographic patterning of evaporated or electroplated layers) packed together and aligned with the incident x rays. On each plate, microstrip charge collectors are focused toward the x-ray source to collect charges generated in the ionization medium (e.g., air or gas) surrounded by high-density materials that act as x-ray converters. The collected charges represent the x-ray image and can be read out by various means, including a two-dimensional (2-D) active readout matrix. The QE, spatial resolution, and sensitivity of the detector have been calculated. It has been shown that the new design will have a QE of more than an order of magnitude higher and a spatial resolution equivalent to that of flat panel systems currently used for portal imaging. The new design is also quantum noise limited down to very low doses (approximately 1-2 radiation pulses of the linear accelerator).
    [Abstract] [Full Text] [Related] [New Search]