  • Title: Simultaneous real-time amperometric measurement of catecholamines and serotonin at carbon fibre 'dident' microelectrodes.
    Author: Pennington JM, Millar J, L Jones CP, Owesson CA, McLaughlin DP, Stamford JA.
    Journal: J Neurosci Methods; 2004 Dec 30; 140(1-2):5-13. PubMed ID: 15589328.
    Abstract:
    Amperometry allows real-time measurement of in vivo electrochemical signals, albeit with no capacity to identify the constituents of the signal. In the present study, differential amperometry was used to monitor catecholamine and serotonin (5-HT) simultaneously at the same location. 'Dident' carbon fibre microelectrodes (microelectrodes with two working electrodes in a single assembly) were constructed and coated with Nafion to prevent poisoning on contact with brain tissue. One electrode (channel A) was held at +200 mV versus Ag/AgCl to monitor catecholamines selectively. This potential is too low to oxidise 5-HT. The second electrode (channel B), recording faradaic current at +500 mV versus Ag/AgCl, measured the sum of catecholamine and 5-HT oxidation. The 5-HT current component was the difference of channel B minus channel A. With appropriate balancing of the two channels, it is possible to record catecholamines and 5-HT simultaneously at the same dident microelectrode. Examples of measurements in striatum, cortex and locus coeruleus are shown.