These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Measurement of breath-by-breath gas exchange during general anaesthesia using a single pneumotachograph.
    Author: Beatty PC.
    Journal: J Biomed Eng; 1992 Mar; 14(2):147-52. PubMed ID: 1564922.
    Abstract:
    A breath-by-breath gas exchange measurement system using a single pneumotachograph suitable for use during general anaesthesia is described. The system's accuracy has been assessed by a combination of error sensitivity analysis, laboratory testing of the component measurements used to calculate gas exchange and measurements on volunteers and patients. The system is shown to have a mean accuracy of +/- 2.6 ml breath-1 for VCO2 measurements, +/- 7.12 ml breath-1 for VO2 and +/- 5.55 ml breath-1 for VN2O measurement. The application of a lung gas stores correction using argon improved between breath variability by 50%.
    [Abstract] [Full Text] [Related] [New Search]