These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Tomotherapy planning of small brain tumours.
    Author: Yartsev S, Kron T, Cozzi L, Fogliata A, Bauman G.
    Journal: Radiother Oncol; 2005 Jan; 74(1):49-52. PubMed ID: 15683669.
    Abstract:
    Helical tomotherapy (HT) combines a rotating intensity modulated fan beam with integrated CT imaging for high precision radiotherapy. HT plans for 12 patients with small brain tumours were compared with five other radiotherapy techniques. Proton techniques gave overall the best results, while HT was shown to produce better target dose uniformity (average SD=1.3%) and kept irradiation of organs at risk as good as other photon methods.
    [Abstract] [Full Text] [Related] [New Search]